Rheonics sensor are shipped factory calibrated. They do not need re-calibration in field, however for various regulatory (ISO, FDA, others) fullfillment of requirements, it may be necessary to verify or re-calibrate the sensors periodically.
This article uses various third party companies and organizations, there use here is by no means an endorsement of any product or company mentioned in this article. Users are responsible for making their own decisions regarding the selection and should review information from the respective supplier to assess the safety risks and proper management of those risks.
SRV measures a product of dynamic viscosity and density.
To compare the measured values against reference viscosity values from viscosity standard, you can follow one of the following procedure.
Preparation
1. Getting suitable reference viscosity standards, we recommend the following companies and standards bodies that provide traceable viscosity standards
- Cannon Instruments is the supplier of NIST traceable viscosity standards, widely available in USA, Europe, Australia, Asia.
- Nippon Grease standards, generally available in Japan
- PTB (Physikalisch-Technische Bundesanstalt) is the national metrological body in Germany. Newtonian calibration liquids for viscosity are supplied by calibration laboratories accredited by Deutsche Akkreditierungsstelle GmbH (DAkkS). These liquids are traceable to the National Standard of Viscosity according to DIN EN ISO/IEC 17025.
2. Sensor handling and preparation. To calibrate or verify the sensor calibration, it is important to ensure the operator follows proper procedure for handling the sensor, calibration fluids and maintain suitable operational conditions.
- Sensor should be placed ideally in a clamp or sensor holder
- Fluid should be poured in a calibration cylinder around the sensor or in a beaker in which the sensor is dipped
- Calibration fluid should not have bubbles
- Let the temperature stabilize to ensure accurate measurements
Next image represents the main considerations for SRV calibration.Next image represents the main considerations for SRD calibration.
Learn more about SRD's thermal equilibrium in this article: Maintaining the SRD's temperature balance for high density accuracy
Review the accessories and setup Rheonics offers to use the SRV and SRD as table top instruments.
Getting Calibration fluid data
Reference fluids are generally provided with measured data by the calibration labs which follow ASTM, ISO, DIN, EN standards. However there is a wide variety of calibration fluids available in the market. Here is a short analysis of the various available fluids and our recommendations.
- Rotation viscometer calibration fluids (silicones): these are typically silicone oils which are provided with a single kinematic value at a single temperature. Like the rotation viscometers (eg Brookfield), these are very inaccurate test fluids and useless for any calibration since the fluid and rotational viscometers have both quite high inaccuracy. Further silicone oils are viscoelastic so their viscosity depends on the frequency of oscillation of instrument. This will give very different results for the silicone oils when measured with a vibrational instrument like the SRV, SRD, DVP.
- Mineral oils: these are Newtonian and ideal for viscometer calibration. However often they only have single measurements. For calibration or verification, it is better to get these oils with both density and viscosity values measured at various temperatures. You will need both density and dynamic viscosity of the oil to perform proper calibration check on the SRV.
- RHEONICS RECOMMENDED - NIST traceable viscosity references: These can be sourced from Cannon Instruments in USA and worldwide. There are other suppliers. We will refer to calibration fluids we source from Cannon Instruments in this document.
NIST Traceable viscosity standards from Cannon Instrument
Viscosity reference selection based on target range
SRV sensor:
- SRV measures a product of dynamic viscosity and density
- SRV has a density output, which is by default set to 1.000
- For SRV, calibration check or re-calibration we recommend using any two of the following: S20, S60, S200, S600
- For higher viscosity ranges, N4000 can be used
- It is important to ensure the higher viscosity fluids (S200, S600, N4000) are poured carefully in the measurement container to avoid entrapping air. Also, heating the fluid to 80 °C and then cooling it down to the required temperature helps remove any entrapped air.
- Rheonics provides sensor holder, calibration cylinders which can be used for calibration, check accessories page
- Rheonics provides a temperature control system, STCM for doing multi-point temperature ramps useful in calibration and testing. Learn more here: STCM info page
Calibration/verification procedure:
SRV density set to 1.0 (default setting).
If you have the density set to a different value, see the section below.
11. Enter the correct density from the calibration reference data
12. Now SRV will output dynamic viscosity
13. Compare reference fluid dynamic viscosity against dynamic viscosity output from SRV
14. Estimate calibration accuracy = measured viscosity by SRV/reference dynamic viscosity of calibration fluid
SRD Sensor: The SRD sensor measures kinematic and dynamic viscosity, so the measurements from the SRD should be compared against the values of the bottle. Once the sensor has stabilize its reading follow these steps:
5. Record SRD Viscosity and density outputs, take 5 readings-your variance should be less than 1% in between these values
6. Compare the values vs the bottle
7. Estimate Calibration accuracy= Measured viscosity by SRD/Dynamic viscosity from the bottle
Measured density by SRD/Density from the bottle
Troubleshooting
What are the common issues?
1. Sensor is not in the fluid completely.
Answer: Make sure sensor sensing element is completely submerged in the fluid. The picture below shows a potential issue since the sensing element is not completely in the fluid.
2. Viscosity value is still changing. Temperature is not stable.
Answer: Wait long enough to stabilize temperature or use an oven or STCM.
3. SRV viscosity output is compared against only dynamic viscosity of the calibration fluids.
Answer: SRV viscosity output should be compared against dynamic viscosity only if the calibration fluid density is entered in the SRV electronics using RCP (see RCP manual for help). If SRV shows a default density output of 1.000 g/cc, then you should compare the viscosity output of SRV against dynamic viscosity multiplied by density of the calibration fluid.
4. How to get density from only dynamic viscosity and kinematic viscosity on the calibration fluid datasheets?
Answer: If you divide dynamic viscosity by kinematic viscosity, you will get density in g/cc.
ISO 17025 certification compliance
For factory calibration and calibration verification Rheonics uses NIST traceable viscosity and density reference fluids from Cannon Instruments. Cannon Instrument holds ISO 17025 and 17034 certification for calibration and production of references. These certificates can be accessed from Cannon Instrument website.
When a calibration certificate is required, ensure it is ordered prior to shipment. CUS option in the calibration is required for the calibration certificate that contains information on the Reference fluid and Lot number.
References:
Cannon Instruments NIST Newtonian viscosity standards >> weblink >> pdf in attachment
PTB Newtonian calibration liquids for viscosity >> weblink >> pdf in attachment
Cannon Instrument ISO 17025 and ISO 17034 certificates >> weblink >> A2LA certificate page >> pdf in attachment