Rheonics sensors output the measurements which are referred as parameter. These start at index 0. The output measurement for each parameter is shown in the list below.


Parameter List: 

Index

Parameter

0

viscosity median / 5 measurements

1

density median / 5 measurements

2

temperature median / 5 measurements

3

Kinematic Viscosity

4

density average / 5 measurements

5

V - viscosity raw

6

D - density raw

7

T - temperature raw

8

f - resonant frequency (Hz)

9

fc - compensated resonant frequency (Hz)

10

df - damping (Hz)

11

Tc - Coil temperature

12

viscosity median and last good

13

density median and last good

14

Displays of mapped value from Modbus register 512

15

Displays of mapped value from Modbus register 514

16

Displays of mapped value from Modbus register 516

17

Te - Estimated Temperature (from internal temperature Algorithm)

18

Tp - Temperature from PT1000 sensor present in sensor

19

Calculated parameter from viscosity model

20

Calculated parameter from density model

21

Calculated parameter from concentration model

22

Sensor cleanliness ratio