Profinet Troubleshooting
	 	
			
			Print
 		
  
	
	Modified on: Mon, 5 Jun, 2023 at 10:46 PM
	
	
	
		| Electronics Issues | 
| No LED lights, display, or output signals | Check power supply and cabling
 | 
| Viscosity is not stable | SRV viscosity output is compared against only the dynamic viscosity of the calibration fluids.The sensor is not fully in fluid.
   | 
| Probe Issues | 
| NaN values are displayed on the RCP software or PLC | Check wiring from the probe to SME is correct.Measure the internal resistance from the probe
   | 
| Communication Issues | 
| No data is visible | Try to reset the SME and/or restart your master.Check all settings are correct with RCP and controller side.Verify there is no duplicated address in the networkTry to use the ping command to get a response from either the PLC or SME.Verify that the latest version of the GSD file is installed.Verify that the station name is unique over the network.
   | 
| Data issues | 
| The measured value is different from my reference standards | SRV viscosity output is compared against only dynamic viscosity of the calibration fluids
 | 
| Wrong Unit from the data stream |  Unit is received as HEX representation, convert to decimal
 | 
| Wrong readings for each parameter | Verify each parameter is correctly mapped, some bytes are swapped
 | 
| Sensor Status does not match any bit from the sensor status table | Sensor status byte is swapped
 | 
	
	
	
		
		
										Did you find it helpful?
								Yes
								No
Send feedback	Sorry we couldn't be helpful. Help us improve this article with your feedback.