What products are involved?
SRV - Inline Viscosity Meter
What is the purpose of this article?
To give an overview of the SRV-X3 probe with Tri-Clamp connection, installation guidelines, instructions for use, and available accessories.
TABLE OF CONTENTS
- 1. Introduction
- 2. General Installation Guidelines
- 3. Process Connection installation instruction
- 4. Accessories available
1. Introduction
SRV is Rheonics sensor for inline viscosity measurement. Read more about SRV here.
Tri-Clamp (TC) is a widely used process connection, especially present in hygienic environments such as industries of food, pharmaceutical, personal care, etc. It is convenient for its ease of use, quick assembly and disassembly, and cleanability.
A TC connection is composed of three main parts: the ferrules, a gasket, and a clamp. Find out how to select the correct gasket here. Pressure and temperature ratings vary for a Tri-Clamp connection, based on the type of clamp used.
At Rheonics, SRV probes with Tri-Clamp connection are typically in stock in sizes 1.5” and 2”. Bigger sizes can be requested and lead time for delivery may increase. The SRV-X3 probe features an intrinsically hygienic design, with optional 3-A and EHEDG certificates.
Note: The SRV-X3 old design is kept in this article for reference and archival purposes. If required, please contact the Rheonics Sales team for more information. See drawing: SRV-X3-15T_old-design.
Figure 1: SRV-X3-B15T 1.5” Tri-Clamp probe dimensions.
Specifications | |
Sensor | |
Order code | SRV-X3-BXXT, XX: size |
Connection type | Tri-Clamp (TC) |
Sizes available | From DN25/32/40 - 1.5" - TC 50.5 up See more. |
Drawings and CAD files | |
Notes:
- Smaller Tri-Clamp connections are available for flush sensor versions. Consider the SRV sensing element shown in Figure 2.
- Check if your required Tri-Clamp norm and size meets the requirements for Rheonics SRV here.
- Find CAD files for various TC sizes here.
2. General Installation Guidelines
This section details the installation requirements and guidelines for the SRV sensor, especially focused on the SRV-X3 variant. Follow the SRV sensor manual for complete instructions of installation.
2.1 Submersion
The SRV has a sensing element shown inside the red section in Figure 2. This section is called the sensing area and must be submerged in the fluid of interest. Notice that a clearance of at least 5mm is recommended between the SRV tip and the pipe wall or any other obstruction.
SRV measures whatever is in contact with its sensing element. Incomplete submersion can lead to inaccurate or systematic errors in viscosity readings. This is a common issue with unfilled pipes due to very low flow rate.
When there is a chance of incomplete pipe filling, consider installing the sensor in a flow cell or section that would ensure the section is full before the fluid exits; this can be done by making the outlet higher than the inlet (Figure 3a and 3b). However, here consideration should be given to create self-draining installations that minimize the chances of sedimentation.

Figure 2: SRV sensing area.
Figure 3a shows an installation with the SRV in the IFC-15T-34N-SRV flow cell, inlet and outlet ports are 3/4” NPT threads.
Figure 3b shows an equivalent alternative with the FET-15T flow elbow tee from Rheonics, which has a shorter 1.5” port for the probe, also available in 2”, 3”, 4”, and other sizes.
Both solutions have the inlet below the outlet port to favor the complete immersion of the sensing element in the fluid before it exits the cell.

Figure 3a: SRV mounted in IFC-15T flow cell. Units: mm [in].

Figure 3b: SRV mounted in FET-15T-15T elbow tee. Units: mm [in].
2.2 Stagnation/dead zone
Stagnations or dead zones must be avoided around the sensing area (Figure 2). A stagnation or dead zone is where fluid transfer is poor, and older fluid may not be fully displaced by newer fluid. This leads to incorrect measurements due to part or all the measurement being influenced by stationary fluid that is no longer representative of the actual process fluid.
Consider that some ferrules may cause dead zones in the gaps created by installing long standpipes. Rheonics recommends the use of short ferrules or its WFT-15T ferrule weldolet.
2.3 Orientation independence
SRV has a symmetrical tip, making it possible to face a flowing fluid in any direction. Its shape enables the fluid to be in contact with the tip without creating recirculation zones.

Figure 4: SRV sensing tip.
2.4 Flow
For Newtonian fluids, the flow rate does not affect the measurements, so SRV should measure the same viscosity value in static and moving states of a fluid at the same temperature.
For non-Newtonian fluids (the vast majority of fluids) the flow rate does matter, and viscosity readings may differ between static and moving conditions.
For processes with non-constant flow rates, the recommendation is to install the SRV in a section of the pipeline with the most consistent flow rate to have steady viscosity value.
Flow rate is also relevant to ensure the full submersion of the SRV sensing element into the fluid. For a process with a low flow rate, pipes may not be full of fluid at all points, so SRV should then be placed in a section where the pipe is usually full of fluid (i.e. after a pump) or consider the use of a flow cell.
In all cases, the SRV’s sensing element should be submerged in the fluid. It’s best to avoid installations with long standpipes (i.e., using long weldolet or tee), since that may lead to poor fluid transfer, resulting in measurements that do not reflect the true state of the fluid or, worse, high noise measurements.
When installing in a standpipe, choose the appropriate insertion length of the sensor by selecting a long insertion probe. This allows the sensing element to clear the stagnation zone and be in the fluid that is of interest for measurement, as in the next figure.

Figure 5: SRV-X5 long insertion probe for long standpipes.
3. Process Connection installation instruction
3.1 Connection steps
A Tri-Clamp connection requires two ferrules or flanges, a seal and a clamp, as shown in the next figure.

Figure 6: Tri-Clamp connection components.
The general steps for a Tri-Clamp connection are:
i. Have a welded ferrule in the pipe or tank.
ii. Place the gasket properly into the seat of the ferrule.
iii. Place the probe Tri-Clamp connection on the gasket and verify it is correctly aligned.
iv. Loosen the clamp so it can fit around the ferrule.
v. Place the clamp and verify the ferrules are within the clamp’s groove.
vi. Start tightening the clamp until a leak-proof connection is made.
Be careful not to over-tighten the clamp; this can damage the gasket and ferrules.
3.2 Specific Process Connection installation instructions
- In line: For inline installations, consider the sensing area (Figure 2) and a minimum clearance of the sensor tip to the pipe inner wall (5 mm). A perpendicular Tri-Clamp installation of the SRV is not possible for most 1”, 1.5” or 2” pipes (for 2”, it depends on the pipe norm and ID). In those cases, the installation is recommended to be parallel or with a flow cell; check accessories FET-15T and IFC-15T-34N-SRV. For bigger lines, the sensor can be mounted perpendicular to the flow; Rheonics recommends the WFT-15T ferrule for such cases. For high-viscosity fluids, especially if they are prone to leaving deposits, a parallel installation is recommended even in lines bigger than 2”.
- In tank: Place the sensor at the bottom, wall, or lid of the tank as long as the sensing element can be fully submerged in the fluid and dead zones are avoided. The WFT-15T ferrule is recommended for these installations. Usually, it is useful if the sensor is placed at a tank height that has always fluid.
- Others: Larger pipelines or tanks often need a longer probe (i.e. jacketed tanks or pipes), in those scenarios, the client should use a long insertion probe - Order code: SRV-X5.
For special applications contact the Rheonics Team.
Note: If there is a hygienic requirement, the SRV-X3 can be ordered with 3-A or EHEDG certification and should be mounted with a hygienic adapter (e.g. flush welding flange like ISO 1127).

Figure 7: SRV installation in line.
Important: Avoid using adapters to install the SRV in a Tri-Clamp port when the probe has a different process connection, e.g. threaded. This is the case for NPT to Tri-Clamp adapters, as showm in Figure 8. Rheonics SRV should be ordered with the correct process connection to directly mount it inline without the need for additional adapters that can affect the correct immersion of the sensing element.

4. Accessories available
4.1 Weldolet - WFT-15T
Weldolet for 1.5” Tri-Clamp sensor connection that ensures maximum submersion of the sensing element. Suitable for the SRV-X3-B15T (1.5”). It can be used for process integration into process lines, mixing vessels, reactors, etc. Review its web page for more information.
Figure 9: WFT-15T ferrule
4.2 Flow cells
a. IFC-15T-34N-SRV
Inline adapter for 1.5” Tri-Clamp sensor with 3/4” NPT inlet and outlet ports. This accessory is recommended for DN5 up to DN25 pipes, ensuring optimal flow over the sensing element and minimal stagnation zones. Review its web page for more information.

Figure 10: IFC-15T-34N-SRV
b. FET-XXT-XXT
Tri-Clamp elbow tee for Rheonics SRV-X3 sensors with the same Tri-Clamp for the inlet and outlet ports. Available for the 1.5” Tri-Clamp sensor (DN25/32/40 DIN 32676), as well as 2”, 3”, 4”, and other sizes. Review its accessory page here.

Figure 11: FET-15T-15T
c. FET-15T-XXX
Tri-Clamp elbow tee for 1.5” Tri-Clamp sensor with welding inlet and outlet ports of 1.5” size. Also available for other line sizes on request. Review its accessory page here.
d. FTP-XXT-XXT
Tri-Clamp tee spool piece for perpendicular installation of the SRV with Tri-Clamp 1.5” or 2” process connection. Available with clamped inlet and outlet ports of 2”, 3” or 4” size. Review its accessory page here.

Figure 12: FTP-20T-30T
References[1]: https://rheonics.com/products/inline-viscometer-srv/
[2]: https://rheonics.com/product-accessories/wft-15t/
[3]: https://rheonics.com/product-accessories/ifc-15t-34n-srv/
[4]: https://rheonics.com/product-accessories/fet-xxt/
[5]: https://rheonics.com/product-accessories/ftp-xxt-xxt/